Dutch          
Industrial Measurement & Control, CEM systems, measuring vans, shelter, analysers and gasconditiongcomponents
Laboratory equipment, color-, ph-, ion selective-, polari-, refracometers
Farmer measurements, NH3, emmision
Reduced Oxygen Reduced Breathing Device, Gas mixing systems
Refractometers, Color measurements and Calcium
Service, Preventive and corrective maintenance, 24/7 contracts
Marine emission monitoring
Home > Products > Gas Analysers > SO₂ UV absorption Analyzer

Filter.

Component







































Show More


Type of measurement









Show More

Principal































Show More

SO₂ UV absorption Analyzer

..Eco-designed, ultra-compact, smart & connected continuous gas analyzer, for emission or process gas monitoring. Perfectly suited to control SO2 removal efficiency in processes using Flue-gas desulfurization FGD / DeSOx systems.

SO₂ UV absorption Analyzer MIR 9000ASD
Eco-designed, ultra-compact, smart & connected continuous gas analyzer, for emission or process gas monitoring. Perfectly suited to control SO2 removal efficiency in processes using Flue-gas desulfurization FGD / DeSOx systems.

Smart analyzer including AMS control functionalities: integrated sampling control, automatic zero and span gas injection, external pump control, system alarms display…

Benefits

  • Superior metrological performances for the measurement of SO2
  • Compatible with any type of drying technology (gas cooler, permeation, dilution…)
  • Environmental-friendly and cost-saving analyzer, with ultra low power consumption
  • Ultra-small and lightweight : 3U, 19’’ Rack
  • Highly accurate, excellent stability economical, easy and reduced maintenance
  • Proactive, user-friendly remote communication
  • Real-time calibration graph, animated synoptic, auto-diagnostic, control and maintenance data screens can be displayed while the instrument is operating
  • Breakthrough mechanical design for weight and power saving as well as thermal insulation & reliability
  • Smart analyzer including AMS control functionalities: integrated sampling control, automatic zero and span gas injection, external pump control, system alarms display…
  • Includes embedded Communication Protocol for WEX® Management Software with automatic recognition and configuration
  • Built-in cell for O2 monitoring (optional)

Main applications
SO2 removal efficiency in processes using Flue-gas desulfurization FGD / DeSOx

Combustion is often a dirty process, releasing a range of pollutants including sulfur dioxide, nitrogen oxides, carbon dioxide, volatile organic compounds, fly ash and a range of heavy metals. Neutralization/absorption of acidic pollutants, namely the NOx and SOx gases, prior to the fly ash filtration and treatment is essential to avoid damage to the filter media and excess of emissions. This operation is realized by the DeNOx (NOx reduction) and the FGD (SOx reduction) systems and is carried out by several techniques, including injecting a quantity of lime, limestone slurry, activated carbon or sodium bicarbonate, proportional to the pollutant concentrations.

In FGD applications, to calculate and precisely adjust this pollutant reduction, on-line monitoring of SO2 (precursor of SO3 and sulfuric acid derivates) must be set on the process upstream of the abatement plant. The advantage of flue gas Process monitoring with ENVEA’s MIR 9000ASD analyzer ensures that the pollutant abatement function is optimized, whilst reducing the overuse of valuable materials in the abatement process, in turn providing cost savings to the plant operator and minimizing pollutant releases. Measuring additionally the water content would additionally allow the detection of a possible leak in the boiler.

Industries concerned:

  • Municipal and Hazardous Waste Incinerators
  • Industrial Boilers and Furnaces
  • Power & Combustion
  • Cement Kilns, Chemical & Petrochemical Plants

Technical Specifications
Measurement range 0-100 / 0-1000 mg/m3 user programmable
0-35 / 0-1500 ppm user programmable
Minimum detectable limit (2σ) 0.2 ppm
Noise (σ) 0.1 ppm
Response time (0-90 %) 5 – 40 sec (programmable)
Zero drift < 1 ppm / 7 days
Span drift < 1% / 7 days
Linearity ± 1 %
Sample flow-rate Approximately 25 liter/hour
Pressure and temperature compensation Automatic
Display TFT LCD color screen, resolution: 800 (RGB) x 480, size: 7 inches
Communication MODBUS/RTU, MODBUS/TCP, MODE4, UDP protocol
Power supply 100~250Vac, 50/60Hz + ground or 24V
Power consumption 50 W
(23 W/h with optional 24 V PS)
Working temperature 0 °C to 35 °C
Memory storage of measurement values Capacity : 1 year based on 15-min intervals
Alarms management Continuous detection and identification of anomalies: temperature, flow rate, electric parameters, programmable measurement threshold
Maintenance tests and diagnostics Direct access through the touch screen, remotely, through the ENVEA ConnectTM app or DAHS software
Output connectivity Ethernet (RJ45), 3 x USB ports, 2 dry contacts outputs included
Optional: 4 analog outputs (0-1 V, 0-10 V, 0-20 mA, 4-20 mA), RS232/RS485
USB port Type A USB socket: 2.0 (3.0 compliant)
Zero/span external SV control Contact connector with screw terminals
Dimensions (LxDxH) 483x606x177 mm ; 19’’ rack, 3U
Weight 9 kg (19.9 lbs)

TÜV Certified Product – Tested AMS Regular Surveillance
0000051692

Emission measurement: “MIR 9000 ASD Option”
Certificate Holder: ENVEA S.A. (Environnement S.A.)
Test Mark Number: 0000051692

  • Tested AMS
  • Regular Surveillance

Documentation

envea_catalogue_cems_emission-monitoring_en envea_catalogue_cems_emission-monitoring_en
envea_mir-9000asd_so2-uv-absorption-analyzer_en envea_mir-9000asd_so2-uv-absorption-analyzer_en
ENVEA_Process_Catalogue_EN ENVEA_Process_Catalogue_EN

Ankersmid Process
Samuel Morsestraat 4
7442 DH Nijverdal
Netherlands


(31)-0548-59 59 69
(31)-0548-62 19 72
info@ankersmid.com

© 2024 - Ankersmid Group