The Drummond Nanoject II Auto-Nanoliter Injector is specifically designed to perform ultra-delicate nanoliter injection procedures into oocytes, embryos, and tissues, etc.
  • Automated microprocessor controlled Microinjection Pipet delivering precise nanoliter volumes accurately
  • Remote control and non-rotating plunger eliminates vibration allowing precise injection minimizing possible cell rupture
  • Selectable Injection Volumes: from 2.3 nanoliters to 69.0 nanoliters in 16 increments

Selectable Operating Speeds
Fill Speed Fast
Slow
46 nL per second
23 nL per second
Injection speed Fast
Slow
46 nL per second
23 nL per second
Empty Speed Fast
Slow
230 nL per second
92 nL per second

  • Supplied with sufficient borosilicate glass capillaries for preparing 400 injection tips, standard softening point 780°C
  • Can be mounted on most micromanipulators Note: Universal Adapter 3-000-024-A supplied with unit
  • Separate Foot Switch available to allow hands-free operation
The Drummond Nanoject II Auto-Nanoliter Injector is specifically designed to perform ultra-delicate nanoliter injection procedures into oocytes, embryos, and tissues, etc. (Zebrafish Larvae Video Condensed | Zebrafish Larvae Video Full)

Each Nanoject II Auto-Nanoliter Injector comes complete with 200 borosilicate capillaries that can be pulled to the required tolerance together with a back filling needle and maintenance kit.

Documentation

Nanoject II Data Sheet Nanoject II Data Sheet
Nanoject Comparison Nanoject Comparison

Ankersmid Process