Dutch          
Industrial Measurement & Control, CEM systems, measuring vans, shelter, analysers and gasconditiongcomponents
Laboratory equipment, color-, ph-, ion selective-, polari-, refracometers
Farmer measurements, NH3, emmision
Reduced Oxygen Reduced Breathing Device, Gas mixing systems
Refractometers, Color measurements and Calcium
Service, Preventive and corrective maintenance, 24/7 contracts
Marine emission monitoring
Home > Brands > Pentol SO3 analyser > SO3 Monitor
HOW MUCH SO3 DO YOU GENERATE?

OPERATING PRINCIPLE
The SO3 or H2SO4 in the gas sample is absorbed as sulphate ions (SO42-) in an aqueous solution of propan-2-ol in water.
The solution is passed through a bed of barium chloranilate. The acid chloranilate ions created are measured in a continuous
flow photometer. By maintaining a constant ratio of flow rates for the gas and the propan-2-ol absorbing solution,
the concentration of acid chloranilate ions can be directly related to the sulphate ion concentration in the absorbing solution,
and hence the SO3 concentration in the gas.

TECHNICAL SPECIFICATION
Range 0 - 200 ppm
Accuracy +/- 5% of reading (in calibrated range)
Lag time Approx. 60 seconds
Response (90% FSD) Approx. 120 seconds (depending on umbilical length)
Solution consumption 0.25 - 2 cm3/h (adjustable as required)
Ambient temperature 0 - 40?C
Module dimensions
Analysis 19? unit, 6U high, 500mm deep (27 kg)
Control 19? unit, 4U high, 500mm deep (13 kg)
Power Requirement 230V 50/60 Hz or 110V 50/60 Hz, 350 W
Probe Length 1 m, 1.2 m, 1.5 m, 2 m
Signal Output 4 .. 20 mA output for Recorder
RS232 to PC connector on request

Documentation


Ankersmid Process
Samuel Morsestraat 4
7442 DH Nijverdal
Netherlands


(31)-0548-59 59 69
(31)-0548-62 19 72
info@ankersmid.com

© 2024 - Ankersmid Group